![Varian F1730308 Instruction Manual Download Page 41](http://html.mh-extra.com/html/varian/f1730308/f1730308_instruction-manual_854741041.webp)
Maintenance
3-3
e.
Thoroughly clean the pump body interior using acetone followed by an alcohol rinse. Dry
with clean, dry, oil-free compressd air or heat gun.
f.
Thoroughly clean the jet assembly using acetone, wipe all surfaces with alcohol, and dry
with clean, dry, oil-free compressed air or heat gun.
g.
Install the jet in the pump body.
h.
Check condition of O-rings and replace, if necessary.
i.
Reinstall the pump in the system.
3-4
DISASSEMBLING AND REASSEMBLING THE COLD CAP
1
To disassemble the cold cap:
a
Remove the female coupling, nut, follower, and the gasket located at the end of the
cold cap water line on the outside of the pump.
b
Remove the nut which secures the cold cap to the jet assembly.
c
Lift out the cold cap.
2
To reassemble the cold cap:
a
Insert the end of the cold cap water line into the cold cap port opening at the side of
the pump before setting the cold cap on top of the jet assembly.
b
Replace the nut which secures the cold cap to the jet assembly. Do not overtighten
the nut.
c
Replace the gasket, the follower, the nut, and the female coupling to the side of
the pump.
Figure 3-1. Heater Element Assembly
Summary of Contents for F1730308
Page 2: ......
Page 4: ......
Page 6: ......
Page 16: ...x ...
Page 18: ......
Page 20: ......
Page 22: ...Installation 1 2 Figure 1 1 Speed and Throughput Curves NHS 35 Diffusion Pump ...
Page 26: ...Installation 1 6 Figure 1 3 Section Drawing NHS 35 Diffusion Pump ...
Page 30: ...Installation 1 10 Figure 1 5 240 400 440V Wiring Diagram NHS 35 Diffusion Pump ...
Page 31: ...Installation 1 11 Figure 1 6 380 415V Wiring Diagram NHS 35 Diffusion Pump ...
Page 32: ...Installation 1 12 Figure 1 7 480V Wiring Diagram NHS 35 Diffusion Pump ...
Page 36: ...Installation 1 16 ...
Page 48: ......