![Varian F1730308 Instruction Manual Download Page 39](http://html.mh-extra.com/html/varian/f1730308/f1730308_instruction-manual_854741039.webp)
3-1
SECTION III
MAINTENANCE
3-1
GENERAL
Diffusion pumps generally require little attention when correctly operated. However, it is advisable to
perform some periodic checks to insure continued trouble-free operation. By simple preventive main-
tenance, costly down-time and cleaning procedures can be avoided. A day-to-day log of pump and
system performance will indicate the condition of the pump and marked variations will show the need
for corrective action.
3-2
PERIODIC INSPECTION
The frequency of inspection will depend on the type of system, its operation and utilization. The maxi-
mum interval between inspections is established on the basis of experience. It is recommended that
the following items be regularly examined.
a.
When the pump is cold, check the condition and level of fluid. These can be inspected by
withdrawing a sample through the drain and by visually checking the level of fluid through the
sight glass respectively. (Use new O-ring gaskets when replacing fill and drain plugs.) Slight
discoloration of the fluid does not affect performance.
b.
Loss of the fluid can be caused by the following:
1.
incorrect air admittance procedures and/or admittance of excessive air or other gas
to a hot pump,
2.
inadequate water cooling
3.
continuous operation at inlet pressures above 10
-3
Torr (.133 Pa)
4.
failure to reinsert the foreline baffle in the pump assembly
c.
When the pump is cold, check that the heaters are bolted snugly to the boiler plate and that
all heater terminal connections are tightly fastened inside the junction box.
WARNING
_____________
High voltages (up to 480V) can kill. Always break the primary circuits to
the power supply before starting to work on the heater and/or its wiring.
Make sure that the pump is at room temperature before attempting ser-
vice. Serious burns can result if this Warning is not heeded.
Depending on system process, poisonous or corrosive compounds may
be present when opening the fill or drain. Always wear appropriate
gloves and clothing and use a self-contained breathing apparatus.
An explosion could occur if the fill or drain is opened when the pump is
running or when it is hot. Check oil level only when the pump is cold.
Summary of Contents for F1730308
Page 2: ......
Page 4: ......
Page 6: ......
Page 16: ...x ...
Page 18: ......
Page 20: ......
Page 22: ...Installation 1 2 Figure 1 1 Speed and Throughput Curves NHS 35 Diffusion Pump ...
Page 26: ...Installation 1 6 Figure 1 3 Section Drawing NHS 35 Diffusion Pump ...
Page 30: ...Installation 1 10 Figure 1 5 240 400 440V Wiring Diagram NHS 35 Diffusion Pump ...
Page 31: ...Installation 1 11 Figure 1 6 380 415V Wiring Diagram NHS 35 Diffusion Pump ...
Page 32: ...Installation 1 12 Figure 1 7 480V Wiring Diagram NHS 35 Diffusion Pump ...
Page 36: ...Installation 1 16 ...
Page 48: ......