www.vareximaging.com
15
1313DXT-I/1515DXT-I/2520DX-I/3030DX-I X-ray Detectors
U
SER
M
ANUAL
8.5
Electrical Rating
Electrical ratings for the detectors.
8.5.1
External Power
shows the power requirements for two different input voltages. The range of input
voltage is 12VDC to 32VDC.
Table 7
External Power Supply Specification (if supplied with the unit)
Detector
XP Power Type
Input Voltage
Input
Frequency
Input
Current
DC Output
1313DXT-I/1515DXT-I
AHM85PS15
100V - 240V
50Hz - 60Hz
1.5A
15V
2520DX-I
AHM85PS15
100V - 240V
50Hz - 60Hz
1.5A
15V
3030DX-I
AHM85PS24
100V - 240V
50Hz - 60Hz
1.0A
+24V/3.54A
Table 8
Electrical Rating (if power supply not provided with the unit)
Detector
Continuous Power Rating
Input Voltage
1313DXT-I/1515DXT-I
11 Watts
12VDC - 32VDC
2520DX-I
12 Watts
12VDC - 32VDC
3030DX-I
24 Watts
21VDC - 30VDC
Table 9
External Power
Detector
Supply Voltage
Power on Surge
Current
Power
1313DXT-I/1515DXT-I
24VDC
2.7A
450mA
10.8 Watts
15VDC
2.8A
690mA
10.4 Watts
2520DX-I
24VDC
2.8A
510mA
12 Watts
15VDC
2.9A
810mA
12.2 Watts
3030DX-I
24VDC
2.8A
1000mA
24 Watts