5. Power and Environmental Specifications
Emissions test
EN/IEC
60601-1-2
test level
Compliance
level
Electromagnetic environment — guidance
Note:
Portable and mobile communications equipment should be separated from the system by no less
than the distances calculated or listed in
on page 27.
Warning:
Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should be used no closer
than 30 cm (12 inches) to any part of the MODEL OR TYPE REFERENCE
of the ME EQUIPMENT or ME SYSTEM, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment
could result.
Table 10: Separation — Equipment not life-supporting
Separation (m) at specified frequencies:
Max output
power (watts)
Separation (m)
150 kHz to 80 MHz
D=(3.5/V1)(√P)
Separation (m)
80 to 800 MHz
D=(3.5/E1)(√P)
Separation (m)
800 MHz to 2.7 GHz
D=(7/E1)(√P)
0.01
0.11667
0.11667
0.23333
0.1
0.36894
0.36894
0.73785
1.
1.1667
1.1667
2.3333
10.
3.6894
3.6894
7.3785
100.
11.667
11.667
23.3333
Guidance and manufacturer’s declaration
Use the following guidance tables for emissions and separation distances:
•
•
•
Emissions, not life-supporting
•
Separation distances, not life-supporting
Table 11: Emissions — Equipment and systems
Emissions test
Compliance
Electromagnetic environment guidance
RF emissions CISPR 11
Group 1
Class A
Nexus DR uses RF energy only for its internal function;
therefore, its RF emissions are very low and unlikely to cause
any interference in nearby electronic equipment.
Harmonics IEC
61000-3-2
Class A
Flicker IEC 61000-3-3
Dmax = 4 %
Nexus DR is suitable for use in all establishments other
than domestic and those directly connected to public low-
voltage power supply network that supplies buildings used for
domestic purposes.
28 | Nexus DR Regulatory and Safety Manual, 736-226-G1, Rev. J
Summary of Contents for nexus DR
Page 1: ...Regulatory and Safety Manual 736 226 G1 Rev J...
Page 2: ...Regulatory and Safety Manual 2 Nexus DR Regulatory and Safety Manual 736 226 G1 Rev J...
Page 13: ...1 Regulatory Information 4 Nexus DR Regulatory and Safety Manual 736 226 G1 Rev J...
Page 17: ...2 Statement of Intended Use 8 Nexus DR Regulatory and Safety Manual 736 226 G1 Rev J...
Page 21: ...3 Cleaning Practices 12 Nexus DR Regulatory and Safety Manual 736 226 G1 Rev J...
Page 44: ...Index 35...
Page 45: ......