Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09
19/24
www.vareximaging.com
Guidance and manufacturer’s declaration – Electromagnetic immunity –
Compliance Test Levels for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
Test
frequency
Band
Service
a)
Modulation
Maximum
power
Distance
I
MMUNITY
TEST LEVEL
(MHz)
(MHz)
(W)
(m)
(V/m)
385
380
– 390
TETRA 400
Pulse modulation
b)
1,8
0,3
27
18 Hz
450
430 – 470
GMRS 460;
FRS 460
FM
± 5 kHz deviation
2
0,3
28
1 kHz sine
710
704 – 787
LTE Band 13,
17
Pulse Modulation
b)
0,2
0,3
9
745
780
217 Hz
810
800 – 960
GMS 800/900;
TETRA 800;
iDEN 820;
CDMA 850;
LTE Band 5
Pulse Modulation
b)
2
0,3
28
870
930
18 Hz
1 720
1 700 – 1
990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse Modulation
b)
2
0,3
28
1 845
1 970
217 Hz
2 450
2 400 – 2
570
Bluetooth;
WLAN
802.11 b/g/n;
RFID 2450;
LTE Band 7
Pulse Modulation
b)
2
0,3
28
217 Hz
5 240
5 100 – 5
800
WLAN 802.11
a/n
Pulse Modulation
b)
0,2
0,3
9
5 500
217 Hz
5 785
a)
For some services, only the uplink frequencies are included.
b)
The carrier is modulated using a 50 % duty cycle square wave signal.