Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09
12/24
www.vareximaging.com
4.
Mains isolation
Figure 1 Means of protection
The AmpMC always needs to be connected to an X-ray system that complies with the required
regulations and standards.
5.
Service, maintenance and cleaning
Refer service to a qualified service technician only.
AmpMC models do not require maintenance and will last during the lifetime of the X-ray system. For
calibration see chapter 7.
In case of malfunction of the AEC system, the AmpMC can be checked according to the described test
procedure §5.4.
Safety precautions
When there is structural damage to the housing or cable of the device, label the device as “out of
order” and have the device repaired prior to further use.
Cleaning
Cleaning with a damp cloth is recommended. Use generally available alcohol-based cleaning agents
and do not soak the device with liquid.
Disinfection
Disinfection, when required, with a damp cloth with Isopropyl alcohol is recommended. Before using a
disinfectant, check at a spot on the bottom of the device if the disinfectant will not damage the plastic
and coated metal surfaces.
Do not soak the device with liquid.
Procedure at defects
- Exchange the extension cable (if present).
- Exchange the AmpMC
Secundary
circuit
AmpMC
SolidStateMC