Vapormed MIGHTY MEDIC Instructions For Use Manual Download Page 11

21 

20 

Mains Operation 

The 

MIGHTY MEDIC Vaporizer can 

also be operated with the power 
adapter, even when the batteries are 
completely discharged. Proceed as 
described in Chapter 5.4. “Charging 
the Battery”, page 18 and then 
switch on the Vaporizer. 

Automatic Switching Off 

The automatic switching off is used 

to save battery capacity so that the 

MIGHTY MEDIC Vaporizer can be 

used as long as possible without 
charging. 

The device switches off automatically 

two minutes after the last press of a 
button or the last application. The 
switching off of the 

MIGHTY MEDIC 

Vaporizer is indicated by a short vibra-

tion. 

A short press of the ON/OFF 
switch during operation can 
reset the automatic switching-

off to two minutes. 

5.5. Heating Up 

The 

MIGHTY MEDIC Vaporizer pro-

vides two temperature levels that are 
set in the factory: The basic tempera-
ture at 180°C (356°F) and the booster 
temperature at 195°C (383°F). 

5. COMMISSIONING AND OPERATION OF THE   

MIGHTY MEDIC VAPORIZER

Switch on the device (heating up the 

MIGHTY MEDIC Vaporizer with insert-

ed Dosing Capsule and Cooling Unit 
mounted is recommended). To do this, 
press the orange ON/OFF switch for 
at least half a second. The Vaporizer 
responds with a short vibration. The 
vibration does not take place if there 
is a fault. 

The display signals that heating is in 

progress by showing the set and ac-
tual temperatures. Set the desired 
temperature (see Chapter 7. “The Use 
of Hemp Flowers (Cannabis Flos)”, 
page  25) using the temperature 
buttons to the left and right of the set 
temperature display. Heating up the 
heat exchanger can take up to two 
minutes, depending on the set tem-
perature and battery charge level. 
Once the set temperature is reached, 
this is signaled by a short double vi-
bration. 
If the basic temperature can no longer 
provide adequate evaporation, the 
booster temperature can be activated 
by means of a double-click on the or-
ange-colored ON/OFF switch. Once 
the booster temperature is reached, 
this is signaled by a short double vi-
bration and a blinking Set Tempera-
ture display. 

5. COMMISSIONING AND OPERATION OF THE   

MIGHTY MEDIC VAPORIZER

5.6. Temperature Setting 

The temperature can be set between 
40°C and 210°C (104°F and 410°F) 
(see Chapter 7. „The Use of Hemp 
Flowers (Cannabis Flos)“, page 
25). 
When you press the plus button, the 
set value is increased, and when you 
press the minus button, the set value 
is decreased. When you briefly tap 
on the plus or minus button, the set 
values are increased or decreased 
respectively in 1-degree increments. 
If you keep the plus or minus button 
pressed, the set values start running 
continuously in the corresponding di-
rection. 
The booster temperature can be ad-
justed in the same way. However, it 
cannot be set lower than the previous 
set temperature. 
The current actual temperature can be 
read from the upper line of the display. 

Fahrenheit/Celsius 
If you simultaneously tap on the plus 
button and the minus button, the dis-
play changes from degree Celsius to 
degree Fahrenheit and vice versa. 

5.7. Reset to Default Settings 
It is possible to reset settings. The 
reset is made by pressing the ON/
OFF button for ten seconds while 
the 

MIGHTY MEDIC Vaporizer is 

switched on. If the ON/OFF button 
has been pressed for ten seconds, 
the signaling of the reset is made by 
vibration. The basic temperature is re-
set to 180°C (356°F) and the booster 
temperature  is  reset  to  195°C  (383°F).            

5.8. Default Settings 
Basic temperature:  

180°C (356°F) 

 Booster  temperature:   
 

additional 15°C (27°F),  

so corresponding to   195°C (383°F) 

Minus temperature button

Plus temperature button

Actual temperature display

Set temperature display 
(displayed blinking:  
booster temperature)

Battery charging display

Summary of Contents for MIGHTY MEDIC

Page 1: ...INSTRUCTIONS FOR USE Read carefully before using the device and save these instuctions...

Page 2: ...GHTY MEDIC Vaporizer 17 5 1 Unpacking 17 5 2 Power Connection 17 5 3 Application and Operation 18 5 4 Charging the Battery 18 5 5 Heating Up 20 5 6 Temperature Setting 21 5 7 Reset to Default Settings...

Page 3: ...zer 36 10 1 Cleaning of the MIGHTY MEDIC Vaporizer and the Power Adapter 36 10 2 Care and Maintenance Storage Disposal 36 10 3 Troubleshooting Periodic Safety Checks 37 10 4 Technical Specifications 3...

Page 4: ...have an average expected operating life of approx 500 operating hours approx 2 years For exchanging the batteries please contact our Service Center Service Contact in case of technical problems or qu...

Page 5: ...2005 The product may not be disposed of in normal household waste The X on the trash can indicates the necessity of disposing of this product separately The device contains a built in rechargeable lit...

Page 6: ...Do not wind the power adapter cord tightly nor pull it over sharp edges and never crush or kink it If the power adapter cord is damaged please contact our Service Center to have it replaced Never try...

Page 7: ...s The Filling Chamber is made of metal do not touch it when it is hot The Dosing Capsule is made of metal do not touch it when it is hot Before cleaning disconnect the power adapter from the socket Th...

Page 8: ...ychotropic mind altering effect Thus there is a potential for abuse which governments in nearly all coun tries of the world counteract with cor responding laws Use only medicinal hemp flowers cannabis...

Page 9: ...e Keep the Instructions for Use for future ref erence Remove the MIGHTY MEDIC Vapor izer and the accessories from the box Immediately after unpacking check whether the Vaporizer the accesso ries the p...

Page 10: ...e frame and all six fields are visible The number of status fields visible depends on the extent of the battery charge Check the housings of the MIGHTY MEDIC Vaporizer and the power adapter for signs...

Page 11: ...on If the basic temperature can no longer provide adequate evaporation the booster temperature can be activated by means of a double click on the or ange colored ON OFF switch Once the booster tempera...

Page 12: ...ules by the nursing staff Inserting the Dosing Capsules into the MIGHTY MEDIC Vaporizer is con siderably easier than filling the Filling Chamber with ground hemp flowers 6 1 Filling the Dosing Capsule...

Page 13: ...or any other cannabis that has not been sub jected to recognised quality controls The amount of active ingredients in these products is generally unknown may fluctuate considerably and they often cont...

Page 14: ...the MIGHTY MEDIC Vapor izer has reached the set temperature 180 C 356 F set and actual tem peratures are the same also signaled by a short double vibration and the Mouthpiece with Lip Piece is folded...

Page 15: ...be used 8 HYGIENE Material Resistance of the MIGHTY MEDIC Vaporizer Please ensure that the Lip Piece the Cooling Unit with Mouthpiece and the MIGHTY MEDIC Vaporizer do not come in contact with the fol...

Page 16: ...or brittle Lip Pieces are not to be used 8 2 Cooling Unit with Mouthpiece When the vapors are cooled it is in evitable that a fraction is condensed on the internal surfaces of the Cool ing Unit The co...

Page 17: ...iece more frequently As an alternative to warm water with wash up liquid all parts of the Cool ing unit with Mouthpiece exception Cap Lock can be cleaned with etha nol Manual cleaning of the disassemb...

Page 18: ...h the upper Screen downwards out of the Filling Chamber Cap from above by using the Filling Chamber Tool supplied Reassemble in reverse order When assembling make sure that Seal Rings are properly sea...

Page 19: ...alified persons In Case of Problems If the power adapter has been in use unplug it at the socket and ensure that no one can connect the MIGHTY MEDIC Vaporizer to the mains supply again without this be...

Page 20: ...o be documented If the device is not functioning prop erly or fails any of the above tests the device has to be repaired 10 4 Technical Specifications Power adapter Manufacturer Mean Well types GSM36E...

Page 21: ...on on Electromagnetic Compatibility Immunity test IEC 60601 test level Compliance level Electromagnetic envi ronment guidance Guidance and Manufacturer s Declaration Electromagnetic Immu nity The MIGH...

Page 22: ...test level 10 THE MIGHTY MEDIC VAPORIZER Immunity test IEC 60601 test level Compliance level Electromagnetic envi ronment guidance Guidance and Manufacturer s Declaration Electromagnetic Immu nity The...

Page 23: ...con sidered If the measured field strength in the location in which the MIGHTY MEDIC Vaporizer is used exceeds the applicable RF compliance level above the MIGHTY MEDIC Vaporizer should be observed t...

Page 24: ...regular mesh width REF 03 11 MM MIGHTY MEDIC Normal Screen Set 6 pcs Lower Screen with regular mesh width REF 03 12 MM MIGHTY MEDIC Coarse Screen Set 6 pcs Upper Screen with coarse mesh width REF 04 0...

Page 25: ...users Normal wear or tear to the products is not con sidered a defect under warranty law If the customer has commissioned third parties with completing maintenance or repair of the product or if the c...

Page 26: ...imated costs and obtained cus tomer approval Copyright This document is copyright protected and may not be used either in whole or in part pursuant to Sections 15 et seqq UrhG German Copyright Act wit...

Page 27: ...AL 30 210 ENG 08 2017 Subject to alterations All rights reserved Manufacturer and Distributor STORZ BICKEL GMBH CO KG In Gruben cker 5 9 78532 Tuttlingen Germany Phone 49 7461 969707 0 Fax 49 7461 969...

Reviews: