![background image](http://html2.mh-extra.com/html/vante/4160/4160_instruction-manual_3422860029.webp)
Vante ® Model 4160
21
2.13.
Troubleshooting
Table 2.1 4160 TPE sealing system troubleshooting guide
Problem
Possible Cause
Solution
Not sealing
No power to sealer
Make sure the unit is plugged in
and the power switch is turned
"ON".
Ensure the power cord is not
damaged.
Blown fuse
Replace the fuse. Refer to Fuse
Replacement on page 26.
Error message
Heater or sensor failure.
Send to Vante for repair.
Bad or incomplete seal
Incorrect settings
Verify the correct material is
selected.
Verify that the temperature, timing,
and cooling settings are appropriate
for the material and sealing
conditions.
Call Vante customer service for
more information.
Very slow sealing
cycle
Air supply is disconnected
Verify that the air supply tube is
connected to an appropriate source.
Uneven cooling
between heating plates
Possible obstruction in air
supply
Look for any kinks in the black air
hose leading to upper heating plate.
If the air supply is clear, contact
Vante customer service for more
instructions.
Summary of Contents for 4160
Page 1: ...Doc No 41600833 01 Rev A1 INSTRUCTION MANUAL TPE TUBE SEALER Model 4160 Generator Controller...
Page 2: ......
Page 4: ...This page intentionally left blank...
Page 15: ...Vante Model 4160 7 Figure 1 4 Sealing head with safety switch...
Page 16: ...Vante Model 4160 8 This page intentionally left blank...
Page 30: ...Vante Model 4160 22 This page intentionally left blank...
Page 32: ...Vante Model 4160 24 This page intentionally left blank...
Page 36: ...Vante Model 4160 28 This page intentionally left blank...
Page 38: ...Vante Model 4160 30 This page intentionally left blank...