DISINFECTION INSTRUCTIONS

1.  Only use a hot, soapy solution or Isoproyl Alcohol to disinfect your hose. Many of the chemicals 

used in typical disinfectants will cause the plastic to deteriorate quickly and break easily. Vaniman 

offers a surface cleaning spray for specifically disinfecting your hose safely. The Surface Disinfec-

tant Spray can be ordered as P/N 3556.

 

NOTE:

 Never use any disinfectant or cleaning solution that contains 

Hydrogen Peroxide

. If you 

are unsure if a particular solution is usable, please contact Vaniman for guidance.

2.  The nozzle can be removed and washed with hot soapy water, Isopropyl Alcohol, cold sterilized or 

Vaniman’s Surface Disinfectant Spray. THE NOZZLE IS NOT AUTOCLAVABLE.

 

FILTER REPLACEMENT INSTRUCTIONS

*** To prevent additional spread of virus or bacteria, it’s advised to contact the CDC for proper handling 

protocol with specific pathogens***

1.  Turn off the unit.
2.  Wearing full PPE (per the CDCs recommendations), set aside a sealable bag for disposing the 

filter(s).

3.  Remove the four thumb screws on the front of unit and set aside for later use. 
4.  Remove the panel to expose the HEPA and pre-filter. 
5.  Remove either the HEPA filter and/or pre-filter by pulling on the tab. Once removed, place filters into 

a sealable bag for disposal. 

6.  Remove any packaging from the replacement HEPA and pre-filter. Carefully align the HEPA filter 

frame with the outer walls of the unit, sit the filter onto the guide ledge, and slide until the filter stops. 

Once the HEPA filter is installed, carefully place the replacement pre-filter on top of the HEPA filter.

6.  Put the front panel back on the unit and secure with thumb screws.
7.  Dispose of the sealed bag containing the used filter(s) per the CDCs recommendations.
8.  Replace HEPA filter when it becomes excessively dirty, or every 6-12 months. 

MAINTENANCE

1.  

Circuit Breaker

 - The unit is protected by a circuit breaker on the left side of the unit.  In the unlikely 

event  there is an overload condition the breaker may trip.  If this happens reset the breaker by 

pushing on the white button until it clicks into place.

2.  

Motor Protection -

 If the motor becomes too hot it will automatically shutoff to prevent damage. To 

reset the thermal gauge, turn off or unplug the unit and allow the motor to cool down. Once it has 

cooled you may turn it back on and continue use.

3. 

Motor Brushes

 - This long life motor does not have replaceable brushes. Replacement motors can 

be ordered as P/N 97041.