VBT-75P S2 USER’S MANUAL REV 1.0
6
Table 2. Functional Descriptions of VBT-75P S2 Controls and Indicators
Item
Number
Panel Markings
Functional Description
1
HIGH VOLTAGE
High voltage cable connector
2
Built-in 2.5" wide thermal printer
3
USB MEM
USB Flash drive port
4
Input power connector with built-in fuse holder and power switch
5
PUSH TO
"ARM"
Arm switch; press and hold during testing.
6
TEST
FAILED
Test failure indicator. This indicator turns on if the test current exceeds the
present current threshold (100, 200, or 300
μ
A).
7
Back-lit LCD screen (20 characters x 4 lines)
8
HIGH VOLTAGE
ENABLED
This indicator turns on when high test voltage is present at the test leads.
9
Rugged membrane keypad
10
GROUND
High voltage return cable connector
11
GROUND
VBT ground stud. Connect ground stud to substation ground using the provided
cable.