REV 2
IRM-5000P USER’S MANUAL
23
NOTE
If you press any other key besides the
[ENTER]
key, you will be
returned to the beginning of step c.
The following screen will be displayed:
Enter the test duration using the numeric keypad and then press the
[ENTER]
key.
Continue to step d
.
d.
The following screen will be displayed:
Select the desired test voltage by pressing the corresponding key (
[1]
,
[2]
,
[3]
,
[4]
, or
[5]
).
e.
The following screen will be displayed:
1.
YES
Press the
[1]
key (
YES
) if you would like to enter the Device Under Test (D.U.T.)
temperature. The following screen will be displayed:
Type the D.U.T. temperature using the numeric keypad. The screen will be
updated as shown:
Press the
[ENTER]
key and
continue to step f
.
ENTER TEMP (0-254 F)
120
°
F
TERMINAL VTG: 1 V
ENTER TEMP (0-254 F)
°
F
TERMINAL VTG: 1 V
ENTER D.U.T. TEMP?
1. YES 2. NO
TERMINAL VTG: 1 V
1. 500V 2. 1000V
3. 2500V 4. 5000V
5. Custom
TERMINAL VTG: 1 V
Test Time (Minutes):
0 Minutes
TERMINAL VTG: 1 V
Summary of Contents for IRM-5000P
Page 25: ...REV 2 IRM 5000P USER S MANUAL 21 Figure 6 Current Plot from a Resistance Test...
Page 36: ...IRM 5000P USER S MANUAL REV 2 32 Figure 10 Step Voltage Test Results Report Printout...
Page 37: ...REV 2 IRM 5000P USER S MANUAL 33 Figure 11 Step Voltage Test Resistance and Current Plots...
Page 42: ...IRM 5000P USER S MANUAL REV 2 38 Figure 13 Dielectric Discharge Test Resistance Plot...
Page 43: ...REV 2 IRM 5000P USER S MANUAL 39 Figure 14 Dielectric Discharge Test Current Plot...
Page 49: ...REV 2 IRM 5000P USER S MANUAL 45 Figure 15 Typical Test Record Directory Printout...