© July 2023 Van Essen Instruments. All rights reserved.
www.vanessen.com
20
Response:
STX
42
SD
YY/MM/DD HH:MM:SS:XXXXX.XXX:ZZZZZ.ZZZ
CC
Data field length is 37 characters per described format
YY/MM/DD HH:MM:SS = Date/time format 17 characters
XXXXX.XXX = value Level (in cmH
2
O with 3 decimals) 9 characters
ZZZZZ.ZZZ = value temperature (in degrees Celsius with 3 decimals) 9 characters
5.4.6
Read Product ID, Name and Firmware Version
Command:
STX
5
PI
none
CC
Response:
STX
25
PI
PP:XXXXXXXXXX:VVVVVV
CC
Data field length is 20 characters per described format
PP = Diver type 2 characters; type is 19 for TD-Diver and Baro-Diver
XXXXXXXXXX = Product name Diver 10 characters
VVVVVV = Firmware version number 6 characters
5.4.7
Read Product Status and Free Memory
Read the logging status and the free memory of the Diver. The logging status is either STARTED,
STOPPED, or FUTURE START. The free memory indicates how many records (time stamp, pressure and
temperature value) can be read until the Diver memory is full.
Command:
STX
5
PS
None
CC
Response:
STX
25
PS
XXXXXXXXXXXXX:MMMMMM
CC
Data field length exactly 20 characters per described format
XXXXXXXXXXXXX = Logging status Diver (STARTED, STOPPED, FUTURE START) 13 characters
Example
Description
22/05/06
␣
16:31:25:
␣
1045.050:
␣␣␣␣
2.374
37 characters (all ASCII)
Example
Description
19:
␣
Diver
␣␣␣␣
:
␣
V1.24
20 characters (all ASCII)