Response:
•
Response will only follow when STX is detected.
•
Response will follow the command with a delay of 0 to 500 ms (depending on OpCode and/or
other Diver operations).
•
If the frame format is incorrect, the OpCode is not supported or the checksum is not correct,
then a communication error will be sent when STX is detected.
List of Commands
5.4.1
Read Date/Time
Read the date/time of the Diver clock.
Command:
STX
5
CL
None
CC
Response:
STX
22
CL
YY/MM/DD HH:MM:SS
CC
Data field length is 17 characters per described format
YY/MM/DD HH:MM:SS = Date/time format 17 characters
5.4.2
Read Monitoring Point Name
Read the name of the monitoring point programmed by the user.
Command:
STX
5
MP None
CC
Response:
STX
25
MP XXXXXXXXXXXXXXXXXXXX
CC
Data field length must be 20 characters
XXXXXXXXXXXXXXXXXXXX = Monitoring point name 20 characters
5.4.3
Read Serial Number
Read the unique serial number of the Diver.
Command:
STX
5
SN
None
CC
Example
Description
MW17-ob
␣␣␣␣␣␣␣␣␣␣␣␣␣
20 characters (all ASCII)