Van Guard 1400 Series Operation Manual Download Page 9

Setup

Note:  If you purchased a 1420BR, 1421BR, 1430BR or a 1431BR this step completes the setup.  Skip ahead to page 

11, “Using Your 1400 Series Microscope.”  If you purchased a 1423PH or a 1433PH, continue on to the next page, 

“Aligning the Phase Contrast Annulus Rings.”

9

1400BR Brightfield Models 

Centering the Condenser

NOTE: For the 1400PH’s please disregard the condenser centering instructions; these models only require phase 

ring alignment, as described on the following page.

Adjustments to the substage condensing system are crucial for proper illumination and performance. There 

are three basic adjustments which need to be made:  Centering, Vertical Focusing, and Aperture Adjustment.

Centering

:  The 

condenser

 must be centered in the light path to ensure proper light control.  A simple method 

for centering is as follows:  

 

  

  Rotate the 

nosepiece

 until the 10X 

objective

 is in the light path.

  

  Raise the substage assembly fully by turning the substage adjustment knob counter-clockwise.

 

 

  Open the 

aperature iris diaphragm

 to the largest setting by using the 

aperature iris diaphragm adjustment

  

lever which extends from the 

condenser

 assembly.

 
  

  While looking into the microscope eyepieces, close the 

field iris diaphragm

 to the smallest setting by turning 

the uppermost section of the 

substage illuminator

 counter-clockwise.

 

  

  Closing the iris in this manner will reduce the field so that a small white hexagon is visible within a black 

field (see figure 8A).  Focusing of the hexagon is performed by turning the 

coarse/fine focus controls

. This white 

hexagon is the light whichis passing through the field iris and should be centered in the black field.  If not, move 

it to the center (see figure 8B) by tightening and/or loosening the 

condenser centering knobs

.

 

  

  Fine tuning can be done by opening the 

field iris diaphragm

 until the white hexagon almost fills the entire 

field (see figure 8C), and then readjusting (see figure 8D). After centering the condenser open the field iris dia

-

phragm slightly wider than the field of view.

 

Vertical Focusing

:  The 

condenser

 can be raised and lowered with the 

substage adjustment

 knob to focus 

the light for optimal illumination.

 

Aperture Adjustment

:  The light path can be adjusted with the 

aperature iris diaphragm adjustment

 lever 

located just below the condenser.  Aperture adjustments are made to induce contrast into a specimen, not to 

adjust light intensity.

1

2

4

3

Figure 8A

Figure 8B

Figure 8C

Figure 8D

Summary of Contents for 1400 Series

Page 1: ...Covering Models 1420BR 1421BR 1430BR 1431BR 1423PH 1433PH Operation Manual 1400 Series...

Page 2: ...ide holder and a chemical resistant finish Motion is controlled by a right hand low position coaxial control and is driven by a rack and pinion system Focusing Movement Coaxial ultra low position coar...

Page 3: ...el 1423PH Binocular Head 1 ea Stand 1 ea Phase Contrast Condenser 1 ea Phase Contrast Centering Telescope 1 ea 10X Eyepiece 2 ea 10X Plan Phase Objective 1 ea 20X Plan Phase Objective 1 ea 40X Plan Ph...

Page 4: ...art Numbers Scale Reticle Call for Part Numbers Stage Micrometer Call for Part Numbers 20X Eyepieces Call for Part Numbers Objectives Achromatic 4X 10X 20X 40X 60X 100X oil Plan Achromatic 4X 10X 20X...

Page 5: ...Port Mechanical Stage Head Assembly Focusing Tension Control Knob Prism Slider Camera Port Dust Cap Phase Annulus Turret Substage Adjustment Knob Diopter Adjustment Mechanism Interpupillary Adjustment...

Page 6: ...6 1400 Series Parts Power Cord Jack 1400 Series Rear View Carrying Handle Power Switch Variable Lighting Control Up Stop Mechanism 1400 Series Lower Left View Substage Adjustment Knob...

Page 7: ...set screw using the included hex wrench see figure 1 NOTE Do not release the head until it is firmly secured with the head set screw Assembly 1 2 3 4 5 Figure 2 Figure 3 7 All Models If you are setti...

Page 8: ...re 6 Make sure that the text NA 1 25 is facing up as you install the condenser If the condenser will not slide freely through the silver ring and you ve checked that the condenser set screw is backed...

Page 9: ...condenser assembly While looking into the microscope eyepieces close the field iris diaphragm to the smallest setting by turning the uppermost section of the substage illuminator counter clockwise Cl...

Page 10: ...contrast centering telescope should resemble rings superimposed on one another see Figure 7A What is actually being viewed are the phase rings Turn the condenser centering knobs which extend from the...

Page 11: ...stage X Y controls located underneath the right hand side of the stage see figure 9 provide easy and accurate positioning of the sample One complete turn of the longitudinal Y control lower half of th...

Page 12: ...t mechanism located on the right eyetube see figure 12 Once complete the microscope is corrected for your vision Interpupillary and Diopter Adjustments Figure 12 Figure 11 1 2 3 The 100X objective whi...

Page 13: ...vert the image to the camera port The prism slider is the silver knob on the right side of the head assembly see figure 13 Note The 1400 Series Microscopes utilize a 70 30 split sliding prism This spl...

Page 14: ...s is blackened pull the fuse out of it s holder and snap a new fuse in You might need to use a screwdriver to lever the fuse out but be careful not to scratch the circuit board Replace the rear cover...

Page 15: ...1 8mm F O V Plan Achromatic 20X 0 40 N A 0 50mm W D 0 9mm F O V 40X 0 65 N A 0 40mm W D 0 45mm F O V 100X 1 25 N A 0 25mm W D 0 18mm F O V Stage Dimensions 140mm x 140mm Stage Motion Right Hand Coaxia...

Page 16: ...ir or replacement of parts by the End User or an unauthorized third party while under warranty 4 Consumable items such as but not limited to bulbs have failed This warranty expressly excludes transpor...

Page 17: ...Printed in the USA No VGMNL081919 1400BR PH...

Reviews: