18
Modbus
Register
Number of
Modbus
Registers
Description
Format
Properties
30201,
30202
2
Diver real-time pressure × 10
SINT32
Read
30203,
30204
2
Diver real-time temperature × 100
SINT32
Read
30205,
30206
2
Diver real-time conductivity × 1000
†
SINT32
Read
30211,
30212
2
Diver real-time pressure
FLOAT
Read
30213,
30214
2
Diver real-time temperature
FLOAT
Read
30215,
30216
2
Diver real-time conductivity
†
FLOAT
Read
30221,
30222
2
Baro real-time pressure × 10
SINT32
Read
30223,
30224
2
Baro real-time temperature × 100
SINT32
Read
30231,
30232
2
Baro real-time pressure
FLOAT
Read
30233,
30234
2
Baro real-time temperature
FLOAT
Read
30241,
30242
1
Number of records recorded by Diver
(0 to 24000, 0 to 48000 records or 0 to
72000 records)
SINT32
Read
30243,
30244
1
Maximum number of records for this
Diver
SINT32
Read
30245
1
Remaining battery capacity Diver
(0 to 100%)
UINT16
Read
†
The type of conductivity is either normal conductivity or specific
conductivity at 25 °C depending on the setting in the CTD-Diver.
6.3
Reading Diver Memory
For the Read input registers – 16 bit registers:
(Registers for reading the Diver records)