MD 850
Basic Functions
Operation
Sealing Operation
Adjust peeling edge width
After loosening the arrest the peeling edge can be adjusted infinitely between 5 mm and 30 mm
by moving the plate.
Insert sterilization package
The drive is switched on automatically.
Removed sealed sterilization package and allow to cool briefly.
Check strength by tearing open seal.
If unsealed points are present increase the sealing temperature. If the paper discolors or the
foil peels off the temperature is too high.
Determine suitable sealing temperature with sealing tests.
Validation – „Seal Check“
Checking of the sealing parameters: temperature, pressing –on pressure and sealing time by means of “SEAL
CHECK”
This test should be performed before the daily working process and should be documented by
filing the print out.
Additional, visual inspection of the sealing seam with Seal Check
Prior to the test, the device must be ready for operation and the temperature must have been reached.
Device is ready for operation
Specified nominal temperature
Has been reached
Activate SEAL CHECK
Introduce sterilisation
packaging, width: 200 mmm
21-12-1998 10:33 T: 180°C F: 100 N t: 2.8 s Miller
4
Peel test
No
Date
Time
Temperature
Pressing-on pressure
Sealing time
Name
SEAL CHECK
Seal check