Valor™ 7000
Series EN-7
3.2 Turning Scale On/Off
To turn the scale on, press and hold the
On/Zero Off
button for 1 second. The scale performs a
display test, momentarily displays the software version, and then enters the active weighing
mode.
To turn the scale off, press and hold the
On/Zero Off
button until OFF is displayed.
3.3 Weigh Mode
This mode is the factory default setting.
1. If needed, press and hold
Mode
until [
wJeIGH
] (Weigh) is displayed.
2. If required, place an empty container on the pan and press
Tare
.
3. Add sample to the pan or container. The display shows the weight of the sample.
3.4 Percent Mode
This mode measures the weight of a sample as a percentage of a reference weight.
1. If required place an empty container on the pan and press
Tare
.
2. Press and hold
Mode
until [
Percnt
] is displayed. [
Clr.ref
]
(clear reference) will then
display. If no reference weight exists, the scale will display [
Put.ref
], proceed to step 5.
3. Press
No
to use the stored reference weight and proceed to step 6.
4. Press
Yes
to establish a new reference. Scale will now display [
Put.ref
].
5. Add the desired reference material to the pan or container. Press
Yes
to store the
reference weight. The display shows 100%.
6. Replace the reference material with the sample material. The display shows the
percentage of the sample compared to reference weight.
7. To clear the stored reference press and hold
Mode
until [
Percnt
] is displayed. Press
Yes
when [
Clr.ref
] is displayed.
Note:
Press
Function
to view the current reference weight.