47
11.4. Aiming beam not correctly centered
•
Press reset on the treatment information window
•
If the aiming beam is still not centered contact Valon
Lasers or local distributor’s service for help.
11.5. Smart wheel not active
•
If smart wheel is not, active try restarting the system.
The smart wheel needs to be connected to the system
be-fore the system starts. Otherwise it doesn’t work
•
If the smart wheel does not work after restarting the sys-
tem, contact Valon Lasers or local distributor’s service
for help.
•
11.6. Foot switch not active
•
Set the system to ”Laser disabled”- mode and enable
the laser again.
•
Check that the foot switch cable is connected to the
laser console
•
Restart the system.
•
If the foot switch still doesn’t work, contact Valon Lasers
Service.
•
11.7. Error messages
In case the program displays an error message:
•
See section 10 for more detailed information about the
error messages.
• If the error is confirmable, try fixing it and press confirm
•
Shut the computer down and restart it.
•
If the error message is still on the screen, contact Valon
Lasers or local distributor’s service and declare the
num-ber of the error message.
11. Troubleshooting