39
8.1. Emergency Stop Button
•
The emergency stop button is situated
on the front panel of the laser console
above the key switch.
•
In case of an emergency you can shut
down the system by pressing the button
down.
•
To continue the normal use the button
needs to be released.
•
Release the button bypulling and turning
it clockwise according to the white ar-
rows.
8.2. Double Control
To increase treatment safety, the laser emission is ensured
with double control: to emit the laser beam the laser must
be enabled and the footswitch must be pressed down.
Through this double control the physician can interrupt the
treatment at any point by releasing the foot switch.
8. Safety