Service Personnel Safety
1-4
RapidVac™ Smoke Evacuator Service Manual
Service Personnel Safety
General
Warning
During initial set up, inspect the smoke evacuator for any damage that may have
been caused during shipping. If damaged, do not use or attempt to repair. Call
Valleylab for service assistance.
The RapidVac Smoke Evacuator filter is a disposable component that captures
potentially hazardous particles. Handle used filters as you would any
biohazardous material. Dispose of used filters according to your local codes and
regulations, and follow your institutions procedures for disposal.
The smoke evacuation filter has a life of approximately 25 hours of use and
should not be used beyond the specified time. If the Replace Filter indicator
illuminates during operation, the smoke evacuator will continue to run until it is
turned off. Turning power off to the smoke evacuator will require replacement of
the filter before subsequent use.
Failure to change the filter accordingly may result in decreased efficiency in
smoke evacuation and contamination of the motor, vacuum pump, and sound
absorbing media within the smoke evacuator.
The RapidVac Smoke Evacuator filter is disposable. Dispose of used filters
according to your local codes and regulations, and follow your institutions
procedures for disposal
Do not operate this device in potentially explosive environments, such as in the
presence of flammable anesthetics.
Refer routine servicing to qualified biomedical technical personnel.
Use only with the power cord provided and always plug into a grounded outlet.
Do not use two- or three-prong adapters with the smoke evacuator power cords.
The power-cord assembly should be checked periodically for damaged insulation
or connectors. Do not use damaged cords.
Use of extension cords may result in fire hazards.
This device is not intended for the evacuation of fluid. If fluid is expected to be
aspirated to the smoke evacuation filter, a fluid collection device must be installed
with the vacuum hose assembly. Failure to install a fluid collection device could
cause filter blockage and electrical damage. Contact Valleylab Service for
additional information.
Do not connect a wet power cord to the wall receptacle.
Connect the smoke evacuator power cord to a properly grounded receptacle.
Plug the power cord directly into the power receptacle without any adapter plugs.
Use of power plug adapters may result in electric shock.