![Vallen spotWave 201 Instruction Manual Download Page 45](http://html1.mh-extra.com/html/vallen/spotwave-201/spotwave-201_instruction-manual_3233499045.webp)
2022-07
file: spotwave_instructions_manual.docx
45 of 50
Hit Feature Extraction Properties
Feature Extraction
Specification
Arrival time resolution 500 ns
Arrival time range
63 bit at a sample rate of 2 MHz (approx.. 146 000 years)
amplitude resolution
Input Range
94 dB
AE
100 dB
AE
134 dB
AE
A
pk
resolution
(approx.)
1.75 µV
3.5 µV
175 µV
detection threshold
resolution
Input Range
94 dB
AE
100 dB
AE
134 dB
AE
A
Thr
resolution
(approx.)
1.75 µV
3.5 µV
175 µV
RMS resolution
Input Range
94 dB
AE
100 dB
AE
134 dB
AE
U
ms
resolution
(approx.)
1.75 µV
3.5 µV
175 µV
Rise time resolution
500 ns
Duration resolution
500 ns
Hit flags
Flag
Description
T
hit time out (Duration > 100 ms)
A
artificially started hit (after a hit time out),
D
long duration mode
Energy resolution
Input range
94 dB
AE
100 dB
AE
134 dB
AE
Eng. Units
1.53 10
-18
V²s
6.13 10
-18
V²s
1.53 10
-14
V²s
Native Units
1.53 10
-4
eu
6.13 10
-5
eu
1.53 eu
Energy unit
1 eu = 10
-14
V²s
Processing Performance
Performance
Specification
Minimum guaranteed
hit rate
100 hits/s with Vallen Systeme’s spotWave Acquisition software
Hit buffer
200 hit data sets in Acquisition Mode
Logging buffer
64 MB, approximately 1.3 million hit data sets