13.3.
Sensor Connector (in)
The SMA socket labelled Sensor is used to connect an AE sensor to it.
Property
Description
Connector type
SMA (IP67), female
Input impedance
16
kΩ || 12
pF (±50 mV, ±100 mV input range); 500 k
Ω || 940 pF (
±5000 mV
input range)
Input range
(
*
)
94 dB
AE
, (±50 mV), 100 dB
AE
, (±100 mV) 134 dB
AE
, (±5000 mV)
(
*
)
The input range is defined by hardware and marked on the device
13.4.
CCT Connector (out)
The CCT connector (Coupling Check Transducer) is used to connect a coupling check transducer to the
spotWave device.
Property
Description
Connector type
SMA (IP67), female
Output range
3.3 V, 20 mA