10
Chapter 3
Specifications
This section lists the operating and accuracy specifications for Valhalla Scientific Model 4021
Digital Ohmmeter. Accuracy is valid at 25°C ± 5°C for a period of one year from the date of
calibration.
Range/Resolution/Test Current/Accuracy
Table 1
Range Resolution
Maximum
Input
Nominal
Test Current
Accuracy
20
Ω
1m
Ω
19.999
Ω
10mA
±0.05% of rdg, ±0.025% of rng
200
Ω
10m
Ω
199.99
Ω
1mA
±0.05% of rdg, ±0.025% of rng
2k
Ω
100m
Ω
1.9999k
Ω
100µA
±0.05% of rdg, ±0.025% of rng
20k
Ω
1
Ω
19.999k
Ω
10µA
±0.05% of rdg, ±0.025% of rng
200k
Ω
10
Ω
199.99k
Ω
1µA
±0.07% of rdg, ±0.03% of rng
2M
Ω
100
Ω
1.9999M
Ω
100nA
±0.1% of rdg, ±0.15% of rng
Temperature
Temperature Coefficient:
±0.002% per °C (from 0°C-15°C and 35°C-50°C)
not
applicable
for
200k
Ω
range and above.
Operating Temperature Range:
0°C to 50°C
Storage Temperature Range:
-10°C to 70°C
General Specifications
Full Scale Test Voltage:
200mV
Current Source Compliance Voltage:
5V
Terminal
Configuration:
4-Wire
Display
Type:
4
½
Digit
LED’s
Overrange/Overload Indication: Display
Flashes
Conversion
Rate:
3
readings/seconds