At the communications established screen select the Calibration option.
Select Site Calibration and required output units
Low Calibration:
At the
lowest expected tide
, Enter the Low Tide Height, either read from a calibrated Tide Board or staffing
gauge
relative to a known datum
.
Press the
Get Reading
button
– TideMaster will sample pressure readings for 30s and record the Low
height/pressure relationship.