To Auto Memorise the Reading

1. 

Press the        key for 4 seconds; the display will show SWT INTV. 

Using the        or        keys, select the desired interval time from 3 

seconds to 255 seconds. Press the       key to exit. 

2. 

Press the        key for 2 seconds to auto memory mode; the       

symbol will appear on display. When storing one reading, the 

‘NO.XXX    ‘ will appear on display for one second to indicate the 

memorised location. 

3. 

Press the        key again to exit. 

To Recall the Memorised Reading

1. 

Press the        key to review the memorised readings. The display 

will show ‘NO.XXX    ‘. 

2. 

Press the        and        keys to scroll through the memorised 

readings. 

3. 

Press        key again to stop viewing the memorised readings. 

4. 

The memorised data can be read by transferring to PC. 

Auto Data Logging Function

Micro SD card 4GB (maximum 99 blocks).

1. 

Press the        key for 4 seconds; the display will show the SET 

INTV. Using the        or       keys, select the desired interval time 

from 3 seconds to 255 seconds. Press the       key again to exit. 

2. 

Press the        key for 4 seconds to begin the continuously logging 

mode; the display will show the     symbol flashing. 

3. 

Press the        key again to exit. 

4. 

The continuous data logging cannot be read from the meter 

directly. Users can read the data by transferring to PC. 

5. 

The continuous data logging cannot be cleared by the meter, only 

the PC. 

Summary of Contents for RE856

Page 1: ...AUSTRALIA 1300 734 253 sales valen com au www valen com au NEW ZEALAND 0800 734 253 sales valen co nz www valen co nz OPERATION MANUAL RE856 BATTERY TESTER...

Page 2: ...ry increasing growth the Lake Cargelligo Head Office moved into a brand new purpose built facility in 2012 In 2015 Regal Electro successfully completed its rst ENLiFEN Energy Storage Optimiser ESO pro...

Page 3: ...done Upbeat Positive Attitude If our employees have a positive attitude and constantly strive to give your best effort eventually we will overcome our immediate problems and nd you are ready for grea...

Page 4: ...nals 6 Screen Display 7 SPECIFICATIONS 8 General Specifications 8 Electrical Specifications 9 OPERATION 10 Preparation 10 Operation 10 Operation 11 Zero adjust Function 11 Pin type Test Lead Zero Adju...

Page 5: ...RE856 OPERATION MANUAL 5 AUS 1300 734 253 NZ 0800 734 253 NAMES FUNCTIONS OF PARTS 1 2 3 6 9 4 5 7 8 10 17 11 13 14 12 15 16...

Page 6: ...emory and auto data logging interval time setting mode press this key again to exit 7 key To turn the comparator function on and off Hold this key for 2 seconds to set the comparator setting mode pres...

Page 7: ...omparator voltage low limit setting Measured battery charge capacity Measured battery discharge capacity Measured battery total capacity The calculated remaining capacity of the battery Shown when the...

Page 8: ...isplay of or results and beeper for warning and fail results Low In High Low High Voltage Resistance Manual and auto data memory 999 sets can be read by meter and downloaded by PC Auto data logging Mi...

Page 9: ...easurement Measuring current frequency 1kHz 30Hz Range Resolution Accuracy 6V 1mV 0 1 readings 6 digits 60V 10mV Voltage measurement Maximum input voltage 60VDC maximum no AC voltage input DO NOT EXCE...

Page 10: ...a high voltage battery before continuing to measure a low voltage battery first short the measurement leads together This will discharge the DC elimination capacitor which is connected across the lead...

Page 11: ...compared with the measurement range Zero adjust Function The zero adjustment function is to zero the range of resistance The reading during zero adjustments would be taken as zero and will be used to...

Page 12: ...or adjustment The zero adjustment s valid for the currently selected range only as long as the power remains on Powering on the meter would reset all zero adjustment values Clip type Test Lead Zero Ad...

Page 13: ...Connect the yellow miniature plug of the DCA current adaptor to the meter DCA CLAMP ADAPTOR jack 2 USE DCA CURRENT ADAPTOR Press the key of the DCA Current Adaptor to turn on the adaptor the POWER LE...

Page 14: ...be The unit then displays with corresponding symbol and sounds a beeper with the WARNING and FAIL cases Comparator Settings 1 Rotate the resistance range switch to any range 2 Press and hold down the...

Page 15: ...3 Press and keys to select the desired comparator number 01 99 The selected comparator number remains in memory even when the power is turned off 4 Press the key again to stop the comparator function...

Page 16: ...y charge discharge remained capacity is Data Memory Function Manual and Auto Data Memory Function 999 sets To clear the Memorised Data When the memory is full the memory number 999 will appear on the...

Page 17: ...splay will show NO XXX 2 Press the and keys to scroll through the memorised readings 3 Press key again to stop viewing the memorised readings 4 The memorised data can be read by transferring to PC Aut...

Page 18: ...her battery fluid leakage possibility When the meter will not be in use for an extended period of time please remove the batteries out of the meter to prevent the possibility of battery fluid leakage...

Page 19: ...User Account Control windows opens asking Do you want to allow the following program from an unknown publisher to make changes to this computer click Yes d e Online Operation 1 Click Run BatCapTester...

Page 20: ...n A dialogue box will open input the file name select a file and then click the button Download Data Download Data from Hard Disk Click the button seventh from the left on the main screen Download Dat...

Page 21: ...crosoft Excel and locate the saved file in Excel format for example test xls The Text Import Wizard then appears Follow the steps from 1 to 3 to complete Data Convert Excel Open a saved data file in t...

Page 22: ...mpling Time PC Sampling Time Meter sampling rate is the rate at which the meter stores readings Click the button fifth from the left on the main menu Meter Sampling Rate Click the sixth from left on t...

Page 23: ...om the right on the main menu Click to set the meter Set Comparator To erase the EEPROM click the button second from the right on the main menu Click to confirm Erase EEPROM To delete the SD card memo...

Reviews: