Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 43
Measurement Unit Change Command Ux
The measurement unit change command is
Ux
.
where
x
=
0, for miles per hour, [mph]
x
=
1, for knots (default), [knot]
x
=
2, for kilometers per hour, [km/h]
x
=
3, for meters per second, [m/s]
The following is an example:
U3
It sets meters per second for wind speed.
NMEA Standard
The standard variable length, comma-separated, MWV wind message is
defined by NMEA 0183 V2.20 as follows:
$WIMWV,<dir>,<ref>,<spd>,<uni>,<sta>*<chk><cr><lf>
where
$WIMWV =
Fixed
text
<dir>
=
Wind angle, 0 to 359 degrees
<ref>
=
Reference; R=Relative, T=True
<spd> =
Wind
speed
<uni>
=
Wind speed units; K = kmph [km/h],
M = mps [m/s], N = kt
<sta>
=
Status; A = Data Valid, V = Invalid Data
* =
Fixed
text
<chk> =
Checksum
(8-bit XOR, excluding $ and *)
<cr>
=
Carriage return code, ASCII 0DH
<lf>
=
Line feed code, ASCII 0AH
NOTE
When the NMEA Standard message format is selected, the sensor
must have a non-zero output interval setting (parameter 5 in the
configuration menu) since no polling command is defined for this
message type.