Setup with 2 temperature probes
1
2
3
5
6
7
4
1
Power switch
2
Probe splitter
3
Latch of the battery cover
4
Instruction label with short version of this setup instruction
5
Probe connectors of the probe splitter
6
Temperature probes
7
Numbering labels for TMP115 probes (included in the probe mounting accessories)
• Two temperature-only probes of following models: HMP110T, HMP115T and
TMP115 (any combination)
• Probe splitter M8 (Vaisala item code CBL210834)
• Optional: probe extension cable(s)
To support operation with two temperature probes, the RFL100
data logger must have firmware version 1.2.0 or higher. Additionally, make sure
the following components of your viewLinc Monitoring System are updated to at
least the following firmware and software levels:
• AP10 firmware version 3.0
• viewLinc 5.0.2
CAUTION!
You must follow the procedure below to set up RFL100 with the
probe splitter. The procedure includes steps to label the probes to make it easy
to identify the channel assignment later. You can also use the info mode of the
RFL100 to check which probe serial number is connected to which channel.
For more information on the info mode, see
RFL100 User Guide (M211861EN)
.
CAUTION!
7
ENGLISH