Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 37
Set Output Interval
Use the
INTV
command to set or display the output interval.
INTV
[
n xxx
]<cr>
where
n
=
time interval = 0 ... 255
xxx
=
time unit = "S", "MIN", or "H"
Example
:
>intv 1 min
Output interval: 1 MIN
>
The shortest output interval (with n = 0) is approximately one second due
to the internal measurement cycle.
Output Reading Once
Use the
SEND
command to output the reading once in STOP mode.
The output format depends on which parameters the transmitter can
output.
SEND
[
aa
]<cr>
where
aa
= Transmitter address (0 ... 99). Must be specified if the
transmitter is in POLL mode, and a line has not been opened
using the
OPEN
command.
Example:
>SEND
T= 25.2 'C aw= 0.299 H2O= 19 ppm RS= 29.9 %
>