Set Serial Line Response Time
With the
SDELAY
command, you can set delay (response time) for user
port RS-485, or view currently set delay value. Value corresponds to tens
of milliseconds (for example 5 = 0.050 s minimum answer delay). The
value can be set between 0...255.
Example:
>sdelay
Serial delay : 0 ? 10
>sdelay
Serial delay : 10 ?
Set Transmitter Address
Addresses are required only for POLL mode (see Set Serial Interface
Mode on page 32). Use the
ADDR
command to input the RS-485
transmitter address.
ADDR [aa]
where
aa
=
address (0 ... 99) (default = 0)
Example:
the transmitter is configured to address 99.
>ADDR
Address : 2 ? 99
>
Format Measurement Output
Use the serial line command
FORM
to change the format or select a
certain quantities for the output commands
SEND
and
R
.
FORM
[
x
]
where
x = Formatter
string