User's Guide _______________________________________________________________________
60 ___________________________________________________________________ M211399EN-F
Clear User Adjustment of RH Measurement
CRH
[
RESET
]<cr>
Example:
>
pass 9000
>
crh reset
OK
Adjust Temperature Measurement
Use the
CT
command to perform an adjustment of the temperature
measurement. You can do a 1-point adjustment or clear the adjustment
information from the HTM10 module. Note that the factory adjustment
will remain intact when user adjustment is cleared.
Show Current T Adjustment
CT
<cr>
Example (showing default temperature offset):
>
pass 9000
>
ct
Temperature offset : 0.000
1-point Adjustment of T Measurement
Place the transmitter in the reference condition and allow it to stabilize
before entering the adjustment.
CT
[
x
]<cr>
where
x = The reference temperature (in degrees Celsius) that the
transmitter should be showing.
Example:
>
pass 9000
>
ct 23
OK