Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 31
Table 7
Advanced Serial Commands
Command Description
ACAL
Calibrate analog output
AMODE
Set/show analog output mode
ASEL [RH/TD]
Set/show analog output quantities and
scaling
ATEST
Test analog output; forces the output ot the
given value
CALINFO
Set/show calibration info
CRH
Calibrate capacitance measurement after
sensor change
CT
Calibrate temperature measurement
FRESTORE
Restore all module settings to factory
defaults
L
Show user adjustment parameters
LI
Set user adjustment parameters
RHCAL
Calibrate humidity measurement
SMODE
[STOP
/RUN/POLL]
Set the serial interface mode
ZERO
Set correction by trimmers to zero
Device Information and Status
Show Device Information
The
?
command outputs a listing of device information.
?
<cr>
Example:
>?
Device Name : HMM100
SW Name : HM100
SW model : VAISALA
SW version : 1.1.10.3031
Serial number :
Calibrated :
Address : 0
Operating mode : 3W
Status : ON
>
If the module is in poll mode, but a connection has not been opened using
the
OPEN
command, issue the
??
command. For a description of the
serial interface modes, see section Set Serial Interface Mode on page 41.
??
<cr>