USER'S GUIDE____________________________________________________________________
30 __________________________________________________________________ M211024EN-B
List of Serial Commands
All commands can be issued either in uppercase or lowercase.
The notation <cr> refers to pressing the carriage return (ENTER) key on
your computer keyboard. Enter a <cr> to clear the command buffer
before starting to enter commands.
Table 6
Basic Serial Commands
Command Description
?
Show module information
??
Show module information (when in poll
mode)
ADDR
[0
... 99]
Set the module address (for POLL mode)
CALCS
Show measured quantities
CLOSE
Close the temporary connection (returns to
POLL mode)
ECHO [ON/OFF]
Set terminal echo
ENV
Set environmental parameters
ERRS
Display active errors
HELP List
available
commands
INTV [
0
... 255
S
/MIN/H]
Set the continuous output interval
(for RUN mode)
OPEN
[0
... 99]
Open a temporary connection to a POLL
mode device
PASS [passcode]
Access advanced commands (requires
passcode)
R
Start the continuous outputting
RESET
Resets the module
RESTORE
Restores the latest saved settings
(overwrites unsaved settings)
S
Stop the continuous outputting
SAVE
Save changed settings to memory
SEND [0 ... 99]
Output the reading once
SERI [baud p d s]
User Port settings (Default: 19200 N 8 1)
baud: 300 ... 115200
STATUS
Show analog output status
SYSTEM
Show module information and build date
VERS
Show software version of the module
XPRES
Set the value for pressure compensations,
temporarily