The first row shows the reading of the probe/transmitter
connected to the port I, the second row shows the reading of the
probe/transmitter connected to the port II. The third row RH/T is
the difference between the readings of the ports. By pressing
BACK
you return to the selection display.
Figure 8
Example of the RH display
Field Checking and Adjustment of
Humidity by Using a Calibrated
Reference Probe
Follow the steps from 1 to 12 and continue as follows:
13.
Ensure that the probes are located in the same conditions.
14.
Wait until the readings have stabilized. If the difference between
the humidity readings is less than 2 % RH, there is no immediate
need for adjustment.
15.
Adjust the transmitter reading to correspond a reading of the
reference probe by pressing
ADJUST
.
16.
Select
To same as RH
I/II
. Press
SELECT.
17.
Confirm by pressing
YES
18.
Press
BACK,
EXIT,
EXIT
to return to the basic display.
1-Point Humidity Adjustment by
Using a Calibrator
You can use Vaisala humidity calibrator HMK15 to achieve the
reference humidities. MI70 indicator is working as a communicator in
the calibration procedure.
Follow the steps from 1 to 12 and continue as follows:
13.
Insert the transmitter probe to the reference humidity