8.8 TMP1 specifications
Table 34 Measurement performance
Property
Description/value
Measurement range
−70 … +180 °C (−94 … +356 °F)
Accuracy at +23 °C (+73.4 °F)
±0.1 °C (±0.18 °F)
Factory calibration uncertainty
±0.1 °C (±0.18 °F) at +23 °C (+73.4 °F)
Sensor
Pt100 RTD Class F0.1 IEC 60751
1)
Defined against calibration reference. Including non-linearity, hysteresis, and repeatability.
2) Exposing temperature sensor to temperatures outside −20 … +150 °C (−4 … +302 °F) may
cause permanent additional deviation of ±0.1 °C (0.18 °F)
3) Defined as ±2 standard deviation limits. Small variations possible; see calibration certificate.
0
0.1
0.2
0.3
0.4
A
cc
ur
ac
y
[+
/-
°
C]
Temperature [
°
C]
-4
0
-3
0
-2
0
-1
0
0
10
20
3
0
4
0
5
0
6
0
70
8
0
9
0
10
0
11
0
12
0
130
140
150
160
170
180
0.5
-70
-60
-50
Figure 43 TMP1 temperature measurement accuracy over full range
Table 35 Operating environment
Property
Description/value
Operating temperature of probe body
−40 … +80 °C (−40 … +176 °F)
Operating temperature of probe head
−70 … +180 °C (−94 … +356 °F)
Operating environment
Suitable for outdoor use
IP rating
IP66 for probe body
EMC compatibility
EN61326-1, industrial environment