HMP240
User's Guide
M210300en
30
Selecting stop bits:
Full duplex/half duplex:
The processor does not allow the following combinations:
•
no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP240 programme will change the number of stop bits to 2
•
even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the programme changes the number of stop bits to 1
NOTE
The serial bus settings become effective only after reset.
5.3.5.3
Setting the transmitter address
Address is used when more than one transmitter is connected to one serial bus;
this way, it is possible to communicate with one transmitter at a time.
•
Select
Addr
in the main menu; the following is displayed:
•
Pressing ENT returns the programme to the main menu.
•
Pressing CL deletes the old address; enter the new address with the
arrow keys.