HMP240
Appendix 7: Re-gaining
M210300en
126
1.2.3 CRI
Changing re-gaining parameters
CRI <cr>
With command CRI the values of the re-gaining parameters can be changed.
Example: changing the re-gaining interval to 480 min and the settling time
after sensor heating to 360 s
>cri
Interval min : 720 ? 480
Duration s : 120 ?
Temp diff : 0.50 ?
Settl time s : 240 ? 360
k % : 95.00 ?
>
NOTE
To activate new parameter settings immediately,
command RESET must be given. Otherwise the
transmitter starts the re-gaining with new parameters
only when the old re-gaining interval is completed.
1.2.4 FST
Displaying re-gaining status
FST nnn <cr>
nnn = ON or OFF
With FST ON the status of the re-gaining is output at the beginning of the
output line.