GMR20 relay output for GM20 series
2
TECHNICAL DATA
contact ratings
max. 50V 0.5 A
default relay setpoints:
range
relay on
relay off
2000 ppm (standard)
1000 ppm
950 ppm
other measurement ranges:
range
relay on
relay off
0...5000 ppm
2500 ppm
2375 ppm
0...10 000 ppm
5000 ppm
4750 ppm
0...20 000 ppm
10 000 ppm
9500 ppm
(setpoints can be changed with the 19222GM Calibration Software Kit)
GUARANTEE
Vaisala issues a guarantee for the material and workmanship of this product
under normal operating conditions for one (1) year from the date of delivery.
Exceptional operating conditions, damage due to careless handling and
misapplication will void the guarantee.