7.6.5 Enable or Disable Analog Output
Use the
AOUT
command to enable or disable the analog output channel.
AOUT
[ON/OFF]
<cr>
Example (turn analog output off):
aout
off
AOUT : OFF
7.7 Serial Line Output Commands
7.7.1 Start Measurement Output
Use the
R
command to start the continuous outputting of measurement values as an ASCII text
string to the serial line. The format of the measurement message is set with the
FORM
command.
R
<cr>
Example (measurement message in default format):
r
Tdf= 17.75 'F Tdfatm= -5.25 'F H2O= 954 ppm
Tdf= 17.23 'F Tdfatm= -5.72 'F H2O= 930 ppm
Tdf= 17.06 'F Tdfatm= -5.88 'F H2O= 922 ppm
Tdf= 16.97 'F Tdfatm= -5.96 'F H2O= 918 ppm
...
Outputting the results continues in intervals issued with the command
INTV
until stopped.
7.7.2 Stop Measurement Output
You can stop the measurement output by pressing the
ESC
key.
7.7.3 Set Output Interval
Use the
INTV
command to change the output interval of the automatically repeating
measurement messages. The measurement messages are repeated in the RUN mode, or after
the
R
command has been given.
INTV
[n xxx]
<cr>
Syntax
Description
n
time interval, range 0 ... 255
xxx
time unit = "S", "MIN", or "H"