2. Open the terminal program and open a connection to the corresponding COM port using
the default settings
19200
,
8
,
N
,
1
, no flow control.
3. Keep the
Enter
key pressed down and connect the other end of the USB cable to the
transmitter. This will cause the transmitter to start in serial mode, using the default serial
settings.
You can now use the transmitter with the terminal program.
To prevent Modbus mode from being restored on the next power-up, select a different serial
mode with the
smode
command.
7.4 List of Serial Commands
All commands can be issued either in uppercase or lowercase. In the command examples, the
keyboard input by the user is in
bold
type.
The notation <cr> refers to pressing the carriage return (
Enter
) key on your computer
keyboard. Enter a <cr> to clear the command buffer before starting to enter commands.
Table 8 List of Serial Commands
Command
Description
<ESC>
Stop the continuous outputting of RUN mode or R command
?
Show device information
??
Show device information (overrides poll mode)
ADDR
[0 ... 255]
Show or set transmitter address
AERR
Set analog output error level
ALARM
Set dew point level for LED alarm
AMODE
Show analog output mode
AOUT
[ON/OFF]
Enable or disable analog output
AOVER
[ON/OFF]
Allow analog outputs to exceed their range 10 %
ASEL
Set analog output parameter and scaling
ATEST
Test analog output
CLOSE
Close the temporary connection and return the device to POLL
mode
ERRS
Display the currently active errors
FILT
[0 ... 1]
Set output filtering level
FORM
[modifier string]
Set output format
FRESTORE
Restore factory settings
HELP
Show list of currently available serial commands
Chapter 7 – Serial Communication
43
Summary of Contents for DRYCAP DMT143
Page 1: ...M211435EN J User Guide Vaisala DRYCAP Dew Point Transmitter DMT143...
Page 85: ...Appendix A Modbus Reference 83...
Page 86: ...DMT143 User Guide M211435EN J 84...
Page 88: ...DMT143 User Guide M211435EN J 86...
Page 89: ......
Page 90: ...www vaisala com...