DM240FA Duct Installation Flange
The transmitter can be installed directly in the process wall through
DM240FA duct installation flange. When the probe is installed
directly on the process wall or pipe, note that a closing valve may be
needed on both sides of the installed probe so that the sensor head can
be removed from the process for calibration or maintenance.
2
3
4
1
0510-035
Figure 18
DM240FA with DMT142
Numbers refer to Figure 18 above:
1 = Measured
gas
2 = DMT142
transmitter
3
=
DM240FA flange (thread G1/2" ISO)
4
=
Recommended additional hole (plugged) for T
d
field check
reference measurement probe (for example, Vaisala DM70)