Chapter 6 __________________________________________________ Calibration and Adjustment
VAISALA ______________________________________________________________________ 115
Feeding Adjustment Information
This information is shown on the device information fields, see section
Information Display on page 57
.
Using Display/Keypad
1.
If you are not in the adjustment menu, press the
ADJ
button on the
motherboard to open the
ADJUSTMENT MENU
.
2.
Select
Adjustment info
, press the right arrow button.
3.
Select
Date
, press the
SET
button. Enter date by using the arrow
buttons. Press the
OK
button.
4.
Select
i
, press the
SET
button. Enter information text including 17
characters at maximum. Use the arrow buttons. Press the
OK
button.
5.
Press the
EXIT
button to return to the basic display.
Using Serial Line
CTEXT
Use the CTEXT command to enter text to the adjustment information
field.
Example:
>ctext Vaisala/MSL
Calibration text : Vaisala/MSL
>
CDATE
Use the CDATE command to enter date to adjustment information field.
Set the adjustment date in format YYYY-MM-DD.
>cdate 2006-06-12
Calibration date : 2006-06-12
>