Dimensions in mm (in inches)
Pressure calculations
Measure the following parameters of the PMB100 barometer module:
- output voltage
V
out
- reference voltage
V
ref
- module temperature
T
m
(°C)
NOTE
External T-sensor is required for temperature compensation purpose
with minimum ±1°C accuracy.
1°C error in temperature measurement causes 0.14 hPa error in
pressure value.