Modbus
registers
Modbus
byte stream
Serial number string (8 characters):
“A1234567”
‘6’
‘4’
‘7’
‘5’
3534h
3736h
n + 2
n + 3
35h
37h
34h
36h
n + 4
n + 6
n + 5
n + 7
32h
41h
33h
31h
36h
34h
37h
35h
Figure 4 Byte order for string fields
6.5 Modbus register addresses for
AQT530
The following register addresses are available in AQT530. The register addresses are:
• R = Read only
• RW = Read/write
• W = Write only
Table 37 Modbus register addresses
Address
RW
Type
Unit
Register
count
Description
0x0000
R
int16
ppb
1
Nitrogen dioxide (NO
2
) concentration in parts
per billion (ppb) with linear correction
• For a value without linear correction, use
register 0x006C
• For µg/m
3
value, use register 0x0066
0x0001
R
int16
ppb
1
Sulfur dioxide (SO
2
) concentration in parts per
billion (ppb) with linear correction
• For a value without linear correction, use
register 0x006D
• For µg/m
3
value, use register 0x0067
Chapter 6 – Modbus interface description
45