30
20901059_EN_VACUU·SELECT_V1.3_280119
Installation and connection
Installation cut‑out (in control cabinet, lab furniture, cable duct)
a
a
Wall thickness
Dimensions (a) for installation cut‑out
1 mm
0.04 in.
111.5 mm x 111.5 mm
4.39 in. x 4.39 in.
2 mm
0.08 in.
112 mm x 112 mm
4.41 in. x 4.41 in.
3 mm
0.12 in.
112.5 mm x 112.5 mm
4.43 in. x 4.43 in.
Depending on the thickness of the wall, appropriate tolerances
should be allowed for the installation cut‑out.
Mount spring clips
Torx T10
Spring clips + screw fittings D3 x 10
20636593
Cut‑out dimensions
Mount spring clips