page 20 of 58
Type
RC 6
Motor protection
self-hold thermal cutout,
manual reset
Rated current during operation
f.)
at:
100-120 V~ 50/60 Hz
230 V~ 50/60 Hz
A
A
5.7 / 5.1
2.7 / 2.6
Degree of protection IEC 529
IP 40
A-weighted emission sound
pressure level
g.)
(uncertainty K
pA
: 3 dB(A))
dB(A)
50
Inlet
small flange KF DN 16
h.)
Outlet
hose nozzle for tubing I.D. 3/8”
(hose nozzle DN 10 mm)
Dimensions L x W x H approx.
in
(mm)
20.1 x 12.0 x 9.1
(510 x 305 x 230)
Weight with oil filling approx.
lbs.
(kg)
53.4
(24.2)
a.) Partial pressure of permanent gases measured at pump inlet.
b.) The total pressure is higher than the partial pressure because the former includes the vapor
pressure of the pump oil and other condensable vapors (e.g., water); the condition of the oil
(cleanliness, content of hydrocarbons with higher partial pressures) is crucial for this value.
c.) The maximum inlet pressure for water vapor, or rather the maximum inlet pressure for vapor,
cannot be specified in accordance with ISO 21360-2 because it cannot be determined for the
HYBRID pump in accordance with this standard. Because the diaphragm pump reduces the
pressure in the oil-sealed part of the RC 6, however, the water vapor tolerance is considerably
higher than that of a conventional oil-sealed rotary-vane pump.
d.) With inlet sealed and without gas ballast.
e.) Oil temperature and pressure in the oil reservoir are the relevant parameters for maximum
vapor inlet pressure and chemical resistance.
f.) During the start-up period of the pump (first 6 minutes after switching-on), the current draw
may be elevated to twice the nominal current draw during normal running conditions.
g.) Measurement according to EN ISO 2151:2004 and EN ISO 3744:1995 at 230V/50Hz and
ultimate vacuum with exhaust tube at outlet.
h.) 120V version: Additional adapter small flange KF DN 16 / hose nozzle DN 10 mm (material: PP)