Installation and connection
20901552_EN-US_PC301x NT V.sel (Ser.Atb_Ha)_V1.1_160822
33
4.4.2 Exhaust connection (OUT)
WARNING
Risk of bursting due to excess pressure in the ex-
haust line.
Unacceptably high pressure in the exhaust line may cause
the vacuum pump to burst or damage seals.
Ø
The exhaust line (outlet, gas outlet) must always be open
and free of pressure.
Ø
Always lay the exhaust hose so that it is descending, or
take measures to prevent condensate return flow into
the vacuum pump.
Ø
Observe the maximum allowable pressures and pressure
differentials.
Connecting the exhaust hose
-> Example
Exhaust connection
at outlet EX
1. Join the rubber sealing ring
(a)
, the hose nozzle
(b)
and the cap
nut
(c)
as pictured and screw this onto the connection.
2. Slide the exhaust hose
(d)
onto the hose nozzle and, if neces-
sary, route the hose to a fume hood. If necessary, fasten the ex-
haust hose, e.g. with a hose clamp.