Repair
-
Maintenance
-
Return
-
Calibration
Every employer (user) is held responsible for the health
and safety of his employees. This also applies to service
personnel performing repair, maintenance, return or cali-
bration.
The
health and safety clearance form
informs the con-
tractor about any possible contamination of the device and
forms the basis for the risk assessment.
In case of devices which have been in contact with
biological substances of risk level 2 contact the VAC-
UUBRAND service absolutely before dispatching the
device. These devices have to be completely disas-
sembled and decontaminated by the user prior to
shipment. Do not return devices which have been
in contact biological substances of risk level 3 or
4.
These devices cannot be checked, maintained or re-
paired. Also decontaminated devices must not returned to
VACUUBRAND due to a residual risk.
The same conditions apply to on-site work.
No repair, maintenance, return or calibration is possi-
ble unless the correctly completed health and safety
clearance form is returned. Devices sent are rejected
if applicable.
Send a completed copy of the
health and
safety clearance form
to us in advance. The declaration
must arrive before the equipment. Enclose a second com-
pleted copy with the product.
Remove all components from the device that are no origi-
nal VACUUBRAND components. VACUUBRAND will not
be responsible for lost or damaged components that are
no original components.
Drain the device completely of fluids and residues.
Decontaminate the device.
Close all openings airtight
especially if using substances hazardous to health.
To expedite repair and to reduce costs, please enclose a
detailed description of the problem and the product’s op-
erating conditions with every product returned.
IMPORTANT