34
20901115_12+16C_VARIO_sel_V1.0_130618
Installation and connection
4.4.3 Venting connection (option)
DANGER
Risk of explosion by venting with air.
Depending on the application, venting can cause explo‑
sive mixtures to form or other hazardous situations to
arise.
>
>
Never vent processes with air which could form an
explosive mixture.
>
>
In the case of flammable substances, use
only inert gas for venting, e. g., nitrogen
(max. 1.2 bar/900 Torr abs.).
The diaphragm pump and controller have
no
direct ventilation
connection. You can connect different venting valves, e. g., the
supplied
VACUU·SELECT
®
Sensor
with integral venting valve.
(b)
The ventilation connection
(b)
for a
VACUU·SELECT
®
Sensor
is described
below.
Alternatively you can use a larger valve,
e. g., a
VB M‑B
(#20674217) for faster
venting.
Vent with ambient air
1
For venting
(b)
with ambient air, nothing needs to be connected to
the sensor
(a)
.
Vent with inert gas – connect venting valve
1
Required connection material:
Hose for hose nozzle, e. g., sili‑
cone tube 4/5 mm
>
Attach hose
(c)
to the connection of venting valve
(b)
and con‑
nect inert gas (max. 1.2 bar/900 Torr, abs.).
5
Venting valve with hose for venting with inert gas
2
.
1 Not applicable to sensors without an integrated venting valve.
2 Avoid overpressure.
(c)
(b)
(a)
(b)
(a)
VACUU·SELECT
®
Sensor with venting
valve
Summary of Contents for MV 10C VARIO select
Page 26: ...26 20901115_EN_Mx10 12 16C_VARIO_sel_V1 0_130618 Product description...
Page 38: ...38 20901115_EN_Mx10 12 16C_VARIO_sel_V1 0_130618 Installation and connection...
Page 86: ...86 20901115_EN_Mx10 12 16C_VARIO_sel_V1 0_130618...
Page 87: ...87 20901115_EN_Mx10 12 16C_VARIO_sel_V1 0_130618...