page 11 of 22

Use and operation

Installing in a vacuum system:

Avoid throttling losses by using connecting pipes with large diameter and keeping
them as short as possible.

Reduce the transmission of vibration and prevent loading due to rigid pipelines.
Insert elastic hoses or flexible elements as couplings between the pump and rigid
pipes.

Use a suitable valve to isolate the pump from the vacuum system if you need to
allow the pump to warm up before pumping condensable vapours or if cleaning the
pump before it is switched off.

Connect the exhaust to a suitable treatment plant to prevent the discharge of dan-
gerous gases and vapours to the surrounding atmosphere. If necessary use a catch-
pot to prevent the drainage of condensate back into the pump and run the exhaust
line downwards.

During operation:

Max. ambient temperature

: 40 °C

Make sure ventilation is adequate if pump is installed in a housing or if ambient
temperature is elevated.

If pump is installed in altitudes of more than 1000 m above mean sea level check
compatibility with applicable safety requirements, e. g. IEC 60034 (motor may over-
heat due to insufficient cooling).

The motor is shut down by a 

thermal cutout

 in the winding.

Attention: Manual reset is necessary. Switch off the pump or isolate the equipment
from mains. Wait approx. five minutes before restarting the pump.

Attention

: In case of 

supply voltage below 100V

, the lock of the cutout might be

restricted and the pump might restart on its own after sufficient cooling down. Take
suitable precautions, if an automatic restart of the pump may lead to a critical
dangerous situation.

Due to the high compression ratio of the pumps, pressure at the outlet port might be
generated being higher than the max. permitted pressure compatible with the mechani-
cal stability of the system.
Do not start the pump if the pressure difference between inlet and outlet ports exceeds
max. 1 bar. Attempts to start the pump at higher difference may cause blockade and
damage of the motor.

Check compatibility with max. permitted pressure at outlet and max. pressure dif-
ference between inlet and outlet ports.

Prevent internal condensation, transfer of liquids or dust. The diaphragm and valves
will be damaged, if liquids are pumped in significant amounts.

A warm up period (approx. 15 min.) is required to ensure that the rated ultimate total
pressure and pumping speed are attained.

Shutdown:

Short-term:

Has the pump been exposed to condensate?

Allow the pump to continue to run at atmospheric pressure for a few minutes.

Has the pump been exposed to media which may damage the pump materials or form

deposits

?

Check and clean pump heads if necessary.

Long-term:

Take measures as described in section short-term shutdown.

Separate pump from the apparatus.

Close manual gas ballast valve (see ”accessoires”).

Close inlet and outlet port (e. g. with transport caps).

Store the pump in dry conditions.

Summary of Contents for ME 8SI

Page 1: ...s strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 696190 999021 08 05 2008 Technology for Vacuum Systems Instructions for use M...

Page 2: ...only to be used and distributed completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 69 61 90 99 90...

Page 3: ...t see Technical data only Failure to connect the motor to ground may result in deadly electrical shock Before switching on the pump check that mains voltage and current conform with the equipment Chec...

Page 4: ...0 6 0 5 V 0 2 1 0 0 1 z H 0 6 0 5 V 0 3 2 0 0 2 A A 2 4 0 4 1 2 0 2 e g a t l o v y l p p u s f o e g n a r d e t t i m r e p x a M h c t i w s n o i t c e l e s e g a t l o v n o i t n e t t A z H 0...

Page 5: ...of operation Prior to use the reference chambers of some of the manometers depending on the model of manometer have to be ventilated Else not ventilating the manometer may lead to a systematic measuri...

Page 6: ...only to be used and distributed completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 69 61 90 99 90...

Page 7: ...nd unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 999021 07 05 2008 page 1 of 22 ME 8 MZ 4 MD 4 MV 2 Diaphragm...

Page 8: ...pump VACUUBRAND diaphragm pumps are the result of many years of experience in construction and practical operation of these pumps combined with the latest developments in material and manufacturing t...

Page 9: ...duct Manual no 999021 07 05 2008 page 3 of 22 Attention Important notes Not permitted Misuse may cause damage Caution Hot surface Isolate equipment from mains Note Contents Safety information 4 Techni...

Page 10: ...l shock The supply cable may be fitted with a moulded European IEC plug or a plug suitable for your local electrical supply If the plug has been removed or has to be removed the cable will contain wir...

Page 11: ...ping substances which may form deposits inside the pump If there is a danger of the formation of deposits in the pump chamber check inlet and outlet of the pump inspect the pump chamber regularly and...

Page 12: ...or capacitors have a typical durability in the range of 10000 to 40000 h depending strongly on the operation conditions like ambient temperature humidity or load Check every capacitor regularly by mea...

Page 13: ...z H 0 6 0 5 d e e p s d a o l o N n i m 1 0 0 8 1 0 0 5 1 0 0 8 1 0 0 5 1 0 0 8 1 0 0 5 1 0 0 8 1 0 0 5 1 t a t n e r r u c d e t a R z H 0 6 0 5 V 0 0 1 z H 0 6 V 0 2 1 z H 0 6 0 5 V 8 0 2 0 0 2 z H...

Page 14: ...a m g a r h p a i d M P F e v l a v M P F 2 V M 4 D M g n i r O M P F e l z z o n e s o h l e e t s s s e l n i a t s e g n a l f l l a m s l e e t s s s e l n i a t s r e c n e l i s E P A P e s o h...

Page 15: ...s strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 999021 07 05 2008 page 9 of 22 ME 8 MZ 4 inlet outlet with silencer pump rati...

Page 16: ...anged It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 999021 07 05 2008 page 10 of 22 MD 4 MV 2 inlet pump rating plate o...

Page 17: ...y Switch off the pump or isolate the equipment from mains Wait approx five minutes before restarting the pump Attention In case of supply voltage below 100V the lock of the cutout might be restricted...

Page 18: ...o systems at different pressure levels stainless steel FFKM leak rate 10 3 mbar x l s for pres sure differences 500 mbar Hose caoutchouc 10 mm ID 686002 Digital vacuum gauge DVR 2 682902 ME 8 Small fl...

Page 19: ...ment with respect to his product Manual no 999021 07 05 2008 page 13 of 22 Assemble control valve first then align and fasten gauge ME 8 Vacuum adjustment device inlet 696840 An additional part is nec...

Page 20: ...o c t o n g n i r g n i r e t n e C e h t n i k a e l r o d e n o i t i s o p m u u c a v r o e n i l e p i p m e t s y s a h t i w p m u p k c e h C t a y l t c e r i d e g u a g m u u c a v k c e h...

Page 21: ...eposits mainte nance should be carried out frequently Regular maintenance will improve the lifetime of the pump and also protect both man and environment Before starting maintenance vent the system is...

Page 22: ...using a spiky or sharp edged tool e g screw driver we recommend to use a rubber mallet or compressed air to be blown carefully into port Remove head cover carefully from housing cover Note po sition o...

Page 23: ...pport disc Check for washers Do not mix the washers from the differ ent heads Make sure that the original number is reassem bled at the individual pump head Smaller number of washers The pump will not...

Page 24: ...re the pres sure at the inlet port A vacuum of less than 110 mbar should be indicated If the reading is higher recheck the pump chamber and make sure that the valves and the diaphragms are correctly s...

Page 25: ...y product returned for repair We submit quotations only on request and always at the customer s expense If an order is given the costs incurred are offset from the costs for repair or from the pur cha...

Page 26: ...for the returned product that neither toxic corrosive biologically active explosive radio active nor contamination dangerous in any way has occurred the product is free of dangerous substances the oi...

Page 27: ...e nous d clarons que le dispositif d sign ci dessus est conforme aux prescriptions de base des directives EU applicables et indiqu s en ci que concerne conception dessin et mod le vendu par nous m mes...

Page 28: ...s obtained under test conditions for specific applications depend on a number of factors beyond our control It is therefore strictly the users responsibility to very carefully check the validity of ap...

Reviews: