![vacuubrand MD 4 VARIO select Instructions For Use Manual Download Page 19](http://html1.mh-extra.com/html/vacuubrand/md-4-vario-select/md-4-vario-select_instructions-for-use-manual_3206759019.webp)
page 19 of 68
20901106_EN_MD4_MV2_VARIO_select_V1.0_140818
Safety during operation
➨
Adopt suitable measures to prevent the release of dan-
gerous, toxic, explosive, corrosive, noxious or pollut-
ing fluids, vapors and gases. To prevent any emission
of such substances from the pump outlet, install an
appropriate collecting and disposal system and take
protective action for pump and environment.
➨
You must take suitable precautions to prevent any for-
mation of explosive mixtures in the pump chamber or
at the outlet of the pump. In case, e.g., of a diaphragm
failure, mechanically generated sparks, hot surfaces
or static electricity may ignite these mixtures. Use inert
gas for venting, if necessary.
➨
Drain appropriately or otherwise remove any poten-
tially explosive mixtures at the outlet of the pump, or
dilute them to non-explosive concentrations.
➨
Never operate this pump if it has a damaged cord or
plug.
+
If the pump is not working properly, has been dropped
or has fallen into water, contact your pump service pro-
vider.
+
Prevent any part of the human body from coming into
contact with vacuum.
+
Make sure that the exhaust pipeline cannot become
blocked.
+
Attention
: At pressures above approximately 810 Torr
(1080 mbar) the pressure reading becomes incorrect.
The display flashes and a warning message is dis
-
played. Release pressure immediately! Risk of burst-
ing!
+
Comply with applicable regulations when disposing of
chemicals. Take into consideration that chemicals may